The theorem on the laboratory bench
Why a generator can appear to double its amplitude, and why a voltmeter becomes part of the circuit the moment you connect it.
- Model an instrument output or input as part of the circuit.
- Explain 50 Ω generator amplitude conventions with a Thévenin source.
- Quantify DMM/scope loading error instead of assuming an ideal measurement.
- Recognize when the lumped resistance model must become ZTH(ω) and Zin(ω).
Can a voltmeter, oscilloscope or signal generator change the quantity it is supposed to measure?
Yes. Measurement is an electrical interaction. A real source has a finite output impedance, and a real instrument has a finite input impedance. Connecting the two creates a new circuit. Thévenin turns this from an annoying laboratory surprise into a predictable calculation.
04.0Prerequisite recap — the three relations you need
This lesson is designed to stand on its own, but it assumes that a source can be represented locally by a Thévenin voltage and an output resistance. If that idea is unfamiliar, chapters 01–03 give the full derivation.
04.1Measurement is not observation from outside the circuit
Suppose an unknown device presents a terminal voltage VTH and output resistance RTH. Before the instrument is connected, the open-circuit terminal voltage is VTH. After connection, the instrument input resistance Rin draws current. The terminal voltage therefore changes.
The phrase “high-impedance input” is therefore incomplete. High compared with what? A 10 MΩ meter is almost an open circuit for a 10 Ω source, but it creates noticeable error for a source resistance of several megohms.
04.2Worked example — why a generator amplitude can appear to double
Many laboratory generators use an output resistance close to 50 Ω and let the user select the load convention used for the displayed amplitude. Consider a generator configured so that a matched 50 Ω load should receive an amplitude of 1.00 V. The following divider model requires an open-circuit source amplitude of 2.00 V.
Figure 4.1 — 50 Ω source driving a matched 50 Ω input
The source model uses a 2.00 V open-circuit amplitude. Equal source and load resistances divide that amplitude by two.
Now replace the 50 Ω load by a 1 MΩ oscilloscope input while leaving the physical source resistance unchanged.
Nothing “amplified” the signal. The generator was already producing almost 2 V at open circuit; the matched 50 Ω load reduced that by a factor of two. A high-impedance input removes most of that loading.
04.3Worked example — a DMM can under-read a high-resistance sensor
Assume a sensor is accurately represented at DC by VTH = 2.50 V and RTH = 1 MΩ. A voltmeter with an assumed input resistance of 10 MΩ is connected.
Figure 4.2 — A voltmeter is part of the circuit once connected
A 2.50 V sensor with 1 MΩ source resistance is measured by an assumed 10 MΩ voltmeter input. The meter becomes the load.
The meter is not “wrong”. It is measuring the voltage of the circuit that exists after the meter has been connected. The error comes from using that loaded voltage as an estimate of the original open-circuit voltage.
04.4Turn an accuracy requirement into an input-impedance specification
A useful engineer does not only calculate the error after choosing an instrument. The calculation can be inverted: start with the maximum acceptable loading error and derive the minimum required input resistance.
Example: a sensor has RTH = 47 kΩ and the loading contribution to voltage error must stay below 0.25%, so ε = 0.0025.
A nominal 10 MΩ input would therefore fail this particular error budget. This is a more useful conclusion than calling 10 MΩ “high impedance”.
Try before opening the answer
A source has RTH = 100 kΩ. What minimum input resistance keeps DC loading error below 1%?
04.5At higher frequency, input resistance is no longer enough
A real oscilloscope or probe input is not purely resistive. A useful first model is an input resistance in parallel with an input capacitance. For a hypothetical instrument with Rin = 1 MΩ and Cin = 20 pF,
Take a source whose output impedance is approximately resistive with RTH = 100 kΩ. At DC, the 1 MΩ input gives a voltage ratio of 0.909. At 100 kHz, the 20 pF capacitance can no longer be ignored.
The same input that seemed “high impedance” at DC now attenuates the signal strongly and introduces phase shift. The correct concept is therefore input impedance as a function of frequency, not one resistance number.
04.6A bench workflow: predict first, connect second
- Identify the source model. Estimate or obtain the source output resistance/impedance and the expected open-circuit amplitude.
- Identify the instrument input. Check whether the input is 50 Ω, high-Z, 1 MΩ, 10 MΩ, or a frequency-dependent probe specification.
- Write the expected loaded value before connecting. Even an order-of-magnitude prediction catches many setup mistakes.
- Check amplitude convention and termination settings. A generator’s “50 Ω” or “High-Z” display setting changes how its displayed amplitude is interpreted; it does not magically remove physical output impedance.
- Check safe limits. Never enable a 50 Ω termination or connect an unknown source without verifying voltage, power and current limits for the actual instrument.
- Record the configuration. Frequency, amplitude convention, termination, probe mode, cable and instrument input belong in reproducible measurement notes.
04.7Troubleshooting: four common measurement mistakes
04.8Independent practice — from calculation to diagnosis
Problem A — direct calculation
A 5.00 V source has RTH = 200 kΩ. It is measured with a 10 MΩ voltmeter. Find the reading and loading error.
Problem B — design from an error budget
A sensor has RTH = 220 kΩ. Specify the minimum input resistance required for loading error below 0.5%.
Problem C — diagnosis before algebra
A generator set for a 1 V matched-load amplitude gives almost 2 V on a high-Z scope, but about 1 V after 50 Ω termination is enabled. Is the generator defective?
If these problems are difficult because the source model itself is unclear, revisit E-002 before moving to transistor loading in chapter 05.
Review E-00204.9What you should now be able to explain
- Why connecting a measuring instrument changes the circuit being measured.
- Why a 50 Ω source can produce almost twice the matched-load amplitude when observed with high impedance.
- How to calculate loading error and convert an error budget into a minimum input resistance.
- Why “10 MΩ input” is incomplete at higher frequency when capacitance matters.
- Why termination, probe mode and cable/instrument configuration must be recorded as part of reproducible evidence.