Passive probes, compensation & loading
Treat a passive voltage probe as a frequency-dependent network, not as a wire. Derive the low-frequency divider ratio, the compensated RC condition that preserves attenuation versus frequency, effective tip resistance/capacitance, under/over-compensation signatures, ground-lead inductance and ringing, and a quantitative method for choosing 1× versus 10× probing from source impedance, signal amplitude and bandwidth requirements.
- Model a passive probe and oscilloscope input as an RC divider rather than an ideal connection.
- Derive the low-frequency attenuation ratio from the probe series resistance and oscilloscope input resistance.
- Derive the compensation condition RpCp=RiCi that makes the divider ratio frequency independent in the first-order model.
- Derive the effective tip input resistance and capacitance of a compensated divider probe.
- Explain why a 10× probe usually reduces capacitive loading while also attenuating the signal seen by the oscilloscope front end.
- Predict under-compensated and over-compensated step responses from unequal RC time constants.
- Model the ground lead as inductance and explain how loop inductance can create ringing/overshoot even with a properly compensated probe.
- Choose probe attenuation by balancing source impedance, loading, signal amplitude, bandwidth and noise rather than using 10× as an unexplained rule.
A passive probe looks mechanically simple, but electrically it is a designed network. Why does a 10× probe often disturb a circuit less than a direct 1× connection, and why must it be compensated before accurate pulse measurements?
The answer comes from an RC divider whose resistive and capacitive ratios must match.
02.1Low-frequency attenuation comes from an ordinary resistor divider
Let the probe contain a series resistance Rp and let the oscilloscope input resistance be Ri. At sufficiently low frequency, ignore capacitances. The channel sees
For a nominal 10× divider,
The oscilloscope or user interface then rescales the displayed voltage by the declared probe factor. If the scope is configured for 1× while a 10× probe is connected, the display will be wrong by a factor of ten even if the probe itself is electrically perfect.
02.2The resistive divider alone fails when input capacitance matters
The oscilloscope input is not purely resistive. Model it as
If the probe were only a series resistor, the divider ratio would change with frequency because Ci progressively shunts the input resistance. High-frequency components would be attenuated differently from low-frequency components, distorting pulse shape.
A compensated passive divider adds a capacitance Cp in parallel with Rp:
Figure 2.1 — A 10× probe is a compensated RC divider
The one question is: what condition makes the divider ratio independent of frequency?
The first-order model contains two RC branches: Rp is in parallel with Cp in the probe, followed by Ri in parallel with Ci at the channel input. Matching the two time constants makes the divider ratio frequency independent in this model.
02.3Derive the compensation condition
The divider transfer is
If the two RC time constants are equal,
then both impedances contain the same factor 1/(1+sτ). It cancels from numerator and denominator:
Thus the first-order divider has the same attenuation at low and high frequency. For a 10× probe with Rp=9Ri, the condition becomes
This is the mathematical meaning of probe compensation.
02.4What impedance does a compensated probe present at its tip?
Under exact first-order compensation, the series combination has
This is equivalent to
For a 10× divider with Rp=9Ri and Cp=Ci/9,
The probe therefore increases the resistive input impedance by ten and, in this simplified model, reduces the effective capacitance by ten. That is why a compensated 10× probe can load a high-impedance/high-frequency node much less than a direct 1× connection.
02.5Worked loading comparison: 1× versus compensated 10×
Use deliberately generic teaching parameters: the 1× channel model is 1 MΩ ∥ 20 pF. A first-order idealized 10× probe therefore presents approximately 10 MΩ ∥ 2 pF at its tip. Let the source resistance be 100 kΩ and f=100 kHz.
For the direct 1× connection:
For the compensated 10× tip loading:
The 10× probe strongly improves the disturbance at the DUT node in this example, but its internal divider also sends only one tenth of that tip voltage to the oscilloscope front end. Loading and measurement noise/dynamic-range consequences must therefore be considered together.
02.6Under-compensation and over-compensation are unequal time constants
Define
If τp<τi, the probe capacitance is too small relative to the input capacitance: high-frequency components are attenuated too strongly and a square-wave edge appears rounded/slow. This is under-compensation.
If τp>τi, high-frequency components are emphasized relative to the desired ratio and the step can show excessive peaking/overshoot. This is over-compensation.
The square-wave compensation signal is useful because a square wave contains a DC/low-frequency level plus many harmonics. A flat top and faithful edge test whether the divider treats those frequency components consistently.
02.7RC compensation is not the whole high-frequency probe model
Real probes also contain cable effects, distributed capacitance, connector inductance and a ground-return geometry. A long ground lead contributes inductance. In the simplest local model,
Fast edge current through Lg can create an additional voltage error. Together with local capacitance it can form a resonant structure. A rough LC frequency scale is
This is an intuition scale, not a complete probe model. It explains why shortening the return connection and minimizing loop area often reduces ringing and pickup on fast edges.
02.81× versus 10× is a trade-off, not a ranking
The correct choice depends on source impedance, signal amplitude, edge speed/frequency content, required accuracy and actual probe/scope specifications.
02.9A practical loading test can reveal what the model missed
A useful empirical check is to observe a node with one probe, store/record the result, and then connect a second similar probe to the same node. If the waveform changes materially, the first probe was already loading the circuit and the doubled load made the effect visible.
This test does not quantify every parasitic by itself, but it is a direct falsification attempt: an ideally non-invasive measurement would not change when another identical high-impedance observer is attached.
02.10Compensation workflow & mastery exercises
- Read the actual probe attenuation, compensation range, input resistance/capacitance and voltage ratings.
- Set or verify the oscilloscope probe factor so displayed amplitude is correctly scaled.
- Connect the probe to the instrument’s specified compensation/reference signal or another appropriate known edge.
- Use the shortest practical reference connection for the compensation check.
- Adjust only the intended compensation element until the step is neither rounded nor excessively peaked.
- Then estimate DUT loading using the probe’s tip impedance over the frequency range of interest.
- For Ri=1 MΩ, what Rp gives a 10× low-frequency divider?
- If Ci=18 pF and Rp=9Ri, what Cp satisfies first-order compensation?
- Under that idealized compensation, what are Rtip and Ctip?
- What qualitative step shape indicates τp<τi?
- Why can a compensated probe still ring on a fast edge?
- A source has high resistance and modest signal amplitude. Give one reason to prefer 10× and one reason 1× might still be attractive.
- If a second identical probe changes the first trace, what conclusion is justified?
- Why is the actual probe specification required even after learning the ideal compensation equations?
Solutions
1. Rp=9 MΩ.
2. Cp=Ci/9=2 pF.
3. Rtip=10 MΩ and Ctip=1.8 pF in the simplified first-order model.
4. Under-compensation: the edge is too slow/rounded because high-frequency components are attenuated too much.
5. Ground-lead inductance, cable/distributed effects and other parasitics remain outside the simple RC compensation model.
6. 10× reduces loading; 1× can deliver a larger signal to the scope front end and improve visibility for very small signals.
7. The probing is measurably invasive at that node; the measurement model must include loading.
8. Real probes include distributed/parasitic effects, finite bandwidth, compensation ranges and safety ratings not contained in the simple two-RC model.
Chapter 03 will move from the analog input path to the acquisition problem: sampling, record length, time base and trigger are different functions and must not be conflated.